10 C.F.R. § 35.2643 Records of periodic spot-checks for remote afterloader units.
Title 10 - Energy
(a) A licensee shall retain a record of each spot-check for remote afterloader units required by §35.643 for 3 years. (b) The record must include, as applicable— (1) The date of the spot-check; (2) The manufacturer's name, model number, and serial number for the remote afterloader unit and source; (3) An assessment of timer accuracy; (4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and (5) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check. (c) A licensee shall retain a copy of the procedures required by §35.643(b) until the licensee no longer possesses the remote afterloader unit.
Title 10: Energy
PART 35—MEDICAL USE OF BYPRODUCT MATERIAL
Subpart L—Records
§ 35.2643 Records of periodic spot-checks for remote afterloader units.