21 C.F.R. Subpart B—Requirements for Specific Medical Devices
Title 21 - Food and Drugs
(a)(1) Informed medical opinion is in agreement that all preparations offered or intended for ophthalmic use, including contact lens solutions, should be sterile. It is further evident that such preparations purport to be of such purity and quality as to be suitable for safe use in the eye. (2) The Food and Drug Administration concludes that all such preparations, if they are not sterile, fall below their professed standard of purity or quality and may be unsafe. In a statement of policy issued on September 1, 1964, the Food and Drug Administration ruled that liquid preparations offered or intended for ophthalmic use that are not sterile may be regarded as adulterated within the meaning of section 501(c) of the Federal Food, Drug, and Cosmetic Act (the act), and, further, may be deemed misbranded within the meaning of section 502(j) of the act. By this regulation, this ruling is applicable to all preparations for ophthalmic use that are regulated as medical devices, i.e., contact lens solutions. By the regulation in §200.50 of this chapter, this ruling is applicable to ophthalmic preparations that are regulated as drugs. (3) The containers shall be sterile at the time of filling and closing, and the container or individual carton shall be so sealed that the contents cannot be used without destroying the seal. The packaging and labeling of these solutions shall also comply with §800.12 on tamper-resistant packaging requirements. (b) Liquid ophthalmic preparations packed in multiple-dose containers should: (1) Contain one or more suitable and harmless substances that will inhibit the growth of microorganisms; or (2) Be so packaged as to volume and type of container and so labeled as to duration of use and with such necessary warnings as to afford adequate protection and minimize the hazard of injury resulting from contamination during use. (c) Eye cups, eye droppers, and other dispensers intended for ophthalmic use should be sterile, and may be regarded as falling below their professed standard of purity or quality if they are not sterile. These articles, which are regulated as medical devices unless packaged with the drugs with which they are to be used, should be packaged so as to maintain sterility until the package is opened and be labeled, on or within the retail package, so as to afford adequate directions and necessary warnings to minimize the hazard of injury resulting from contamination during use. [47 FR 50455, Nov. 5, 1982] (a) General. Unless contact lens solutions used, for example, to clean, disinfect, wet, lubricate, rinse, soak, or store contact lenses and salt tablets or other dosage forms to be used to make any such solutions are packaged in tamper-resistant retail packages, there is the opportunity for the malicious adulteration of these products with risks both to individuals who unknowingly purchase adulterated products and with loss of consumer confidence in the security of the packages of over-the-counter (OTC) health care products. The Food and Drug Administration has the authority and responsibility under the Federal Food, Drug, and Cosmetic Act (the act) to establish a uniform national standard for tamper-resistant packaging of those OTC products vulnerable to malicious adulteration that will improve the security of OTC packaging and help assure the safety and effectiveness of the products contained therein. A contact lens solution or tablet or other dosage form to be used to make such a solution for retail sale that is not packaged in a tamper-resistant package and labeled in accordance with this section is adulterated under section 501 of the act or misbranded under section 502 of the act, or both. (b) Requirement for tamper-resistant package. Each manufacturer and packer who packages for retail sale a product regulated as a medical device that is a solution intended for use with contact lenses, e.g., for cleaning, disinfecting, wetting, lubricating, rinsing, soaking, or storing contact lenses or tablets or other dosage forms to be used to make any such solution shall package the product in a tamper-resistant package, if this product is accessible to the public while held for sale. A tamper-resistant package is one having an indicator or barrier to entry which, if breached or missing, can reasonably be expected to provide visible evidence to consumers that tampering has occurred. To reduce the likelihood of substitution of a tamper-resistant feature after tampering, the indicator or barrier to entry is required to be distinctive by design or by the use of an identifying characteristic (e.g., a pattern, name, registered trademark, logo, or picture). For purposes of this section, the term “distinctive by design” means the package cannot be duplicated with commonly available material or through commonly available processes. A tamper-resistant package may involve an immediate-container and closure system or secondary-container or carton system or any combination of systems intended to provide a visual indication of package integrity. The tamper-resistant feature shall be designed to and shall remain intact when handled in a reasonable manner during manufacture, distribution, and retail display. (c) Labeling. Each retail package of a product covered by this section is required to bear a statement that is prominently placed so that consumers are alerted to the tamper-resistant feature of the package. The labeling statement is also required to be so placed that it will be unaffected if the tamper-resistant feature of the package is breached or missing. If the tamper-resistant feature chosen to meet the requirement in paragraph (b) of this section is one that uses an identifying characteristic, that characteristic is required to be referred to in the labeling statement. For example, the labeling statement on a bottle with a shrink band could say “For your protection, this bottle has an imprinted seal around the neck.” (d) Requests for exemptions from packaging and labeling requirements. A manufacturer or packer may request an exemption from the packaging and labeling requirements of this section. A request for an exemption is required to be submitted in the form of a citizen petition under §10.30 of this chapter and should be clearly identified on the envelope as a “Request for Exemption from Tamper-resistant Rule.” A petition for an exemption from a requirement of this section is required to contain the same kind of information about the product as is specified for OTC drugs in §211.132(d) of this chapter. This information collection requirement has been approved by the Office of Management and Budget under number 0910–0150. (e) Products subject to approved premarket approval applications. Holders of approved premarket approval applications for products subject to this section are required to submit supplements to provide for changes in packaging to comply with the requirement of paragraph (b) of this section unless these changes do not affect the composition of the container, the torque (tightness) of the container, or the composition of the closure component in contact with the contents (cap liner or innerseal) as these features are described in the approved premarket approval application. Any supplemental premarket approval application under this paragraph is required to include data sufficient to show that these changes do not adversely affect the product. (f) Effective date. Each product subject to this section is required to comply with the requirements of this section on the dates listed below except to the extent that a product's manufacturer or packer has obtained an exemption from a packaging or labeling requirement: (1) Initial effective date for packaging requirements. (i) The packaging requirement in paragraph (b) of this section is effective on February 7, 1983 for each contact lens solution packaged for retail sale on or after that date, except for the requirement in paragraph (b) of this section for a distinctive indicator or barrier to entry. (ii) The packaging requirement in paragraph (b) of this section is effective on May 5, 1983 for each tablet that is to be used to make a contact lens solution and that is packaged for retail sale on or after that date. (2) Initial effective date for labeling requirements. The requirement in paragraph (b) of this section that the indicator or barrier to entry be distinctive by design and the requirement in paragraph (c) of this section for a labeling statement are effective on May 5, 1983 for each product subject to this section packaged for retail sale on or after that date, except that the requirement for a specific label reference to any identifying characteristic is effective on February 6, 1984 for each affected product subject to this section packaged for retail sale on or after that date. (3) Retail level effective date. The tamper-resistant packaging requirement of paragraph (b) of this section is effective on February 6, 1984 for each product subject to this section that is held for sale at retail level on or after that date that was packaged for retail sale before May 5, 1983. This does not include the requirement in paragraph (b) of this section that the indicator or barrier to entry be distinctive by design. Products packaged for retail sale after May 5, 1983, are required to be in compliance with all aspects of the regulations without regard to the retail level effective date. [47 FR 50455, Nov. 5, 1982; 48 FR 1706, Jan. 14, 1983, as amended at 48 FR 16666, Apr. 19, 1983; 48 FR 37625, Aug. 19, 1983; 53 FR 11252, Apr. 6, 1988] Effective Date Note: A document published at 48 FR 41579, Sept. 16, 1983, stayed the effective date of §800.12(f)(3) until further notice. (a) Purpose. The prevalence of human immunodeficiency virus (HIV), which causes acquired immune deficiency syndrome (AIDS), and its risk of transmission in the health care context, have caused the Food and Drug Administration (FDA) to look more closely at the quality control of barrier devices, such as surgeons' gloves and patient examination gloves (collectively known as medical gloves) to reduce the risk of transmission of HIV and other blood-borne infectious diseases. The Centers for Disease Control (CDC) recommend that health care workers wear medical gloves to reduce the risk of transmission of HIV and other blood-borne infectious deseases. The CDC recommends that health care workers wear medical gloves when touching blood or other body fluids, mucous membranes, or nonintact skin of all patients; when handling items or surfaces soiled with blood or other body fluids; and when performing venipuncture and other vascular access procedures. Among other things, CDC's recommendation that health care providers wear medical gloves demonstrates the proposition that devices labeled as medical gloves purport to be and are represented to be effective barriers against the transmission of blood- and fluid-borne pathogens. Therefore, FDA, through this regulation, is defining adulteration for patient examination and surgeons' gloves as a means of assuring safe and effective devices. (1) For a description of a patient examination glove, see §880.6250. Finger cots, however, are excluded from the test method and sample plans in paragraphs (b) and (c) of this section. (2) For a description of a surgeons' glove, see §878.4460 of this chapter. (b) Test method. For the purposes of this regulation, FDA's analysis of gloves for leaks will be conducted by a water leak method, using 1,000 milliliters (mL) of water. Each medical glove will be analyzed independently. When packaged as pairs, each glove is considered separately, and both gloves will be analyzed. A defect on one of the gloves is counted as one defect; a defect in both gloves is counted as two defects. Defects are defined as leaks, tears, mold, embedded foreigh objects, etc. A leak is defined as the appearance of water on the outside of the glove. This emergence of water from the glove constitutes a watertight barrier failure. Leaks within 1 and (1) The following materials are required for testing: A 2 (2) The following methodology is used: Examine the sample and identify code/ lot number, size, and brand as appropriate. Examine gloves for defects as follows: carefully remove the glove from the wrapper, box, etc., visually examining each glove for defects. Visual defects in the top 1 (3) Immediately after adding the water, examine the glove for water leaks. Do not squeeze the glove; use only minimal manipulation to spread the fingers to check for leaks. Water drops may be blotted to confirm leaking. If the glove does not leak immediately, keep the glove/filling tube assembly upright and hang the assembly vertically from the horizontal rod, using the wire hook on the open end of the fill tube (do not support the filled glove while transferring). Make a second observation for leaks 2 minutes after addition of the water to the glove. Use only minimal manipulation of the fingers to check for leaks. Record the number of defective gloves. (c) Sample plans. FDA will collect samples from lots of gloves to perform the test for defects described in paragraph (b) of this section in accordance with FDA's sampling inspection plans which are based on the tables of MIL-STD-105E (the military sampling standard, “Sampling Procedures and Tables for Inspection by Attributes,” May 10, 1989). Based on the acceptable quality levels found in this standard, FDA has defined adulteration as follows: 2.5 or higher for surgeons' gloves and 4.0 or higher for patient examination gloves at a general inspection level II. FDA will use single normal sampling for lots of 1,200 gloves or less and multiple normal sampling for all larger lots. For convenience, the sample plans (sample size and accept/reject numbers) are shown in the following tables: (d) Lots of gloves which are tested and rejected using the test method according to paragraph (b) of this section, are adulterated within the meaning of section 501(c) of the Federal Food, Drug, and Cosmetic Act, and are subject to regulatory action, such as detention of imported products and seizure of domestic products. [55 FR 51256, Dec. 12, 1990]
Title 21: Food and Drugs
PART 800—GENERAL
Subpart B—Requirements for Specific Medical Devices
§ 800.10 Contact lens solutions; sterility.
§ 800.12 Contact lens solutions and tablets; tamper-resistant packaging.
§ 800.20 Patient examination gloves and surgeons' gloves; sample plans and test method for leakage defects; adulteration.
Adulteration Level at 2.5 for Surgeons' Gloves---------------------------------------------------------------------------------------------------------------- Number defective Lot size Sample Sample Number --------------------- size examined Accept Reject----------------------------------------------------------------------------------------------------------------35,001 and above........................ First..................... 125 125 2 9 Second.................... 125 250 7 14 Third..................... 125 375 13 19 Fourth.................... 125 500 19 25 Fifth..................... 125 625 25 29 Sixth..................... 125 750 31 33 Seventh................... 125 875 37 3835,000 to 10,001........................ First..................... 80 80 1 7 Second.................... 80 160 4 10 Third..................... 80 240 8 13 Fourth.................... 80 320 12 17 Fifth..................... 80 400 17 20 Sixth..................... 80 480 21 23 Seventh................... 80 560 25 2610,000 to 3,201......................... First..................... 50 50 0 5 Second.................... 50 100 3 8 Third..................... 50 150 6 10 Fourth.................... 50 200 8 13 Fifth..................... 50 250 11 15 Sixth..................... 50 300 14 17 Seventh................... 50 350 18 193,200 to 1,201.......................... First..................... 32 32 0 4 Second.................... 32 64 1 6 Third..................... 32 96 3 8 Fourth.................... 32 128 5 10 Fifth..................... 32 160 7 11 Sixth..................... 32 192 10 12 Seventh................... 32 224 13 141,200 to 501............................ Single sample............. ......... 80 5 6500 to 281.............................. Single sample............. ......... 50 3 4280 to 151.............................. Single sample............. ......... 32 2 3150 to 51............................... Single sample............. ......... 20 1 250 to 0................................. Single sample............. ......... 5 0 1----------------------------------------------------------------------------------------------------------------
Adulteration Level at 4.0 for Patient Examination Gloves---------------------------------------------------------------------------------------------------------------- Number defective Lot size Sample Sample Number --------------------- size examined Accept Reject----------------------------------------------------------------------------------------------------------------10,001 and above........................ First..................... 80 80 2 9 Second.................... 80 160 7 14 Third..................... 80 240 13 19 Fourth.................... 80 320 19 25 Fifth..................... 80 400 25 29 Sixth..................... 80 480 31 33 Seventh................... 80 560 37 3810,000 to 3,201......................... First..................... 50 50 1 7 Second.................... 50 100 4 10 Third..................... 50 150 8 13 Fourth.................... 50 200 12 17 Fifth..................... 50 250 17 20 Sixth..................... 50 300 21 23 Seventh................... 50 350 25 263,200 to 1,201.......................... First..................... 32 32 0 5 Second.................... 32 64 3 8 Third..................... 32 96 6 10 Fourth.................... 32 128 8 13 Fifth..................... 32 160 11 15 Sixth..................... 32 192 14 17 Seventh................... 32 224 18 191,200 to 501............................ Single sample............. ......... 80 7 8500 to 281.............................. Single sample............. ......... 50 5 6280 to 151.............................. Single sample............. ......... 32 3 4150 to 91............................... Single sample............. ......... 20 2 390 to 26................................ Single sample............. ......... 13 1 225 to 0................................. Single sample............. ......... 3 0 1----------------------------------------------------------------------------------------------------------------