42 C.F.R. § 84.255   Requirements for end-of-service-life indicator.


Title 42 - Public Health


Title 42: Public Health
PART 84—APPROVAL OF RESPIRATORY PROTECTIVE DEVICES
Subpart N—Special Use Respirators

Browse Previous |  Browse Next

§ 84.255   Requirements for end-of-service-life indicator.

(a) Each canister or cartridge submitted for testing and approval in accordance with §§84.252, 84.253, and 84.254 shall be equipped with a canister or cartridge end-of-service-life indicator which shows a satisfactory indicator change or other obvious warning before 1 ppm vinyl chloride penetration occurs. The indicator shall show such change or afford such warning at 80 ±10 percent of the total service life to 1 ppm leakage, as determined by continuing each test described in §§84.252(b), 84.253(b), and 84.254(b) until a 1 ppm leakage of vinyl chloride occurs.

(b) The applicant shall provide sufficient pretest data to verify the performance of the end-of-service-life indicator required in paragraph (a) of this section.