42 C.F.R. § 493.1269   Standard: Hematology.


Title 42 - Public Health


Title 42: Public Health
PART 493—LABORATORY REQUIREMENTS
Subpart K—Quality System for Nonwaived Testing
Analytic Systems

Browse Previous |  Browse Next

§ 493.1269   Standard: Hematology.

(a) For manual cell counts performed using a hemocytometer—

(1) One control material must be tested each 8 hours of operation; and

(2) Patient specimens and control materials must be tested in duplicate.

(b) For all nonmanual coagulation test systems, the laboratory must include two levels of control material each 8 hours of operation and each time a reagent is changed.

(c) For manual coagulation tests—

(1) Each individual performing tests must test two levels of control materials before testing patient samples and each time a reagent is changed; and

(2) Patient specimens and control materials must be tested in duplicate.

(d) The laboratory must document all control procedures performed, as specified in this section.